Optimum X-ray collimation is a must to reduce radiation exposure and to obtain the best image. We started to research and develop it since founding and got success on market.
INNOVA X-Ray Collimator
Overview
Optimum X-ray collimation is a must to reduce radiation exposure and to obtain the best image. We started to research and develop it since founding and got success on market. We own both of manual control collimator and auto control collimator intellectual property rights, Including of hardware and software.
INNOVA C20
A manually operated collimator with large knobs for direct and accurate adjustment of the light field, quick collimation every time, reliable and easy to use.
INNOVA C80
A motorized collimator for integration with radiography and fluoroscopy systems. The collimator is controlled with the front knobs or remotely through its CAN-bus &RS232. A large touch-screen LCD display shows the X-ray tube tilting angle, selected beam filtration and field size with a given SID.
Features
INNOVA C20
- This collimator is standard equipped with a long-life power LED that projects a bright light field on the exposed area.
- The lightcan be customized to be switched on with a push-button at the front or be switched on with foot switch.
- Durable LED positioning lamp, enable to be turned on automatically when you move floating table, and automatically be turned off after 30 seconds.
INNOVA C80
- It equipped with a long-life powerLED that projects a bright light field on the exposed area.
- Aligning the detector under theX-ray beam is made easy with a center laser line.
- The beam filtration is possible with the optionalmotorized filter module. No filtration or three different aluminum plus copper combination filtrations are available.
- Shutter calibrationand customization of user settings are easily performed through the software menu, accessible via the display.
Technical Specifications
Stationary X-ray equipment for radiography is up to 150 kV tube voltage.
- Rectangular Field
INNOVA C20 43 x 43 cm at 1m SID (maximum)
INNOVA C80 48 x 48 cm at 1 m SID (maximum)
1.2 mm Al.eq at 75kVp
- Attenuation equivalent(IEC 60601-1-3:2008 §3.8)
- Light Source
- Power LED, white, adjustable luminosity
LED can be controlled by external
Signal: “on/off” or using internal timer. - Light field illuminance(IEC 60601-2-54:2009 §203.8.102.5)
>160 lx with LED power set to ≈70% (1m from focus, 35 x 35 cm)
- Leakage radiation(IEC 60601-1-3:2008 §12.4)
≤0.5 mGy/hr (at 1m, 150kVp, 4mA)
- Accessory rails
For two (2) accessory slides of 177mm wide x 2mm thickness
- Power ConsumptionAC/DC models
INNOVA C20 12 – 45 VDC or 20 – 30 VAC
INNOVA C80 24 VDC, 35 VA max.
- Additional Filters(IEC 60601-1-3:2008 §7.3 / §7.4)
INNOVA C80
2 additional motor driven filters, manually selectable or by CAN-bus Selectable filter combinations
• No added filter
• Filter 1 =0.1 mm Cu
• Filter 2 = 0.2 mm Cu
• Filter 3 = 0.3 mm Cu
- Bucky center line laser(IEC 60825-1)
INNOVA C80 Class 1 laser
- Weight
INNOVA C20 6.0 kg
INNOVA C80 9.0 kg